New _best_ — Narishige Pc10 Manual
Adjust the heater knob to a value (typically between 4-7, depending on glass type).
Ideal for patch-clamp pipettes requiring a rapid taper and a stiff, low-resistance tip. Configure the toggle switch to .
Connect the unit to a suitable power source, ensuring the voltage matches the device specifications. narishige pc10 manual new
For detailed technical troubleshooting, you can refer to the Narishige FAQ page or download the full PC-10 PDF manual weight and heat settings
Known to last 20+ years with occasional filament changes. The new manual emphasizes cleaning the rails monthly. Adjust the heater knob to a value (typically
: Set the "No.2 Heater" level. This stage finalizes the tip diameter and shape. 4. Error Messages & Troubleshooting Common error codes for the PC-10 include: : Slider distance set to 0mm. : Set to 1mm or longer. : First pull completion not detected. : Increase No.1 Heater output. : Slider speed too fast during first pull. : Reduce No.1 Heater output or weight. : Heater contact failure or broken heater. : Tighten fixing screws or replace heater.
Set the Heater Dial roughly 5% to 10% above your established Melting Reference Point . Connect the unit to a suitable power source,
The story of the is one of a legendary laboratory workhorse that served as the global standard for micropipette fabrication for decades before its retirement. Known for its "gravity-fall" vertical pulling method, it allowed researchers to create ultra-fine needles for microinjection and patch-clamp experiments with remarkable consistency. The Legacy of the PC-10
| Symptom | Possible Cause | Solution | | :--- | :--- | :--- | | | Temperature too low | Increase Heater Knob setting. | | | Heater coil broken | Inspect coil for continuity; replace if necessary. | | Tip is clogged / closed | Temperature too high / Pull too slow | Reduce heat to create a sharper break. | | | Dirty glass | Clean glass with alcohol or acetone before pulling. | | Tip is too wide/blunt | Pull was too fast (too hot) | Lower the heater temperature. | | | Weight too heavy | Use a lighter weight for a longer, finer taper. | | Tip is too long/whippy | Pull was too slow (too cool) | Increase heater temperature. | | | Weight too light | Use a heavier weight. | | Uneven tips | Glass not centered in coil | Adjust the glass holder to ensure the glass is perfectly vertical and centered. |
The Narishige PC-10 is a dual-stage vertical glass micropipette puller widely recognized for its reliability in producing patch-clamp electrodes, microinjection needles, and microelectrodes. Utilizing a gravity-driven pull mechanism and a dual-stage heating system, the PC-10 allows researchers to customize pipette taper length, tip diameter, and shape.